Introduction Pseudohypoxaemia is commonly encountered when interpreting oxygenation in diseases with hyperleucocytosis and thrombocytosis. Various methods of overcoming this problem have been reported. We suggest precooling of the syringes before drawing the blood for analysis as another useful method of reducing oxygen consumption by the hyper metabolic cells. Case Report 41 year old male, diagnosed with chronic myeloid leukaemia (CML) for five years with multiple admissions for easy fatiguability, abdominal pain, distension, fever and vomiting was treated conservatively with hydroxyurea and multiple blood transfusions. On this admission he was posted for splenectomy due to massive splenomegaly. Bilateral basal crepitations was noted on chest auscultation. Investigations showed haemoglobin- 8.7 g.dL-1, total leucocyte (WBC) count- 5, 09, 600 cells mm-3, platelet count- 1,14,000 cells mm-3, renal function, serum electrolytes and ECG were normal. Echocardiogram revealed minimal pericardial effusion. Chest X-ray showed cardiomegaly with normal lung fields. Patient underwent splenectomy uneventfully. Intraoperatively, he maintained oxygen saturation (SpO2) of 99-100%. Intraoperative arterial blood gas (ABG) analysis done on FiO20.5 showed a pH- 7.303, PaCO2- 47.9 mmHg, PaO2- 47.8 mmHg, standard bicarbonate- 23.2 mmol. L-1. In view of poor oxygenation as per the report, he was shifted to intensive care unit (ICU) for ventilator support suspecting pulmonary atelectasis. ABGs done in ICU also showed poor oxygenation [Table 1].
alysis done on FiO20.5 showed a pH- 7.303, PaCO2- 47.9 mmHg, PaO2- 47.8 mmHg, standard bicarbonate- 23.2 mmol. L-1. In view of poor oxygenation as per the report, he was shifted to intensive care unit (ICU) for ventilator support suspecting pulmonary atelectasis. ABGs done in ICU also showed poor oxygenation [Table 1]. Table 1 Serial post-operative ABGs and the corresponding cell counts Based on clinical improvement and pulse oximetry findings of 99-100%, he was weaned of ventilator support, extubated and discharged out of ICU. Postoperatively, patient was diagnosed to be in blast crisis in view of increasing platelet and WBC counts. He was started on hydroxyurea and allopurinol tablets to achieve cytoreduction. He was intubated and readmitted to ICU following an episode of generalised tonic-clonic seizure, altered sensorium and desaturation on oxygen therapy of 60% venturi by face mask. ABGs done after this admission also showed poor oxygenation with acceptable acid base status [Table 2]. Table 2 Serial post-readmission ABGs with the corresponding cell counts
Postoperatively, patient was diagnosed to be in blast crisis in view of increasing platelet and WBC counts. He was started on hydroxyurea and allopurinol tablets to achieve cytoreduction. He was intubated and readmitted to ICU following an episode of generalised tonic-clonic seizure, altered sensorium and desaturation on oxygen therapy of 60% venturi by face mask. ABGs done after this admission also showed poor oxygenation with acceptable acid base status [Table 2]. Table 2 Serial post-readmission ABGs with the corresponding cell counts Chest radiograph did not show any deterioration. Meanwhile patient regained consciousness over a period of one day and tolerated pressure support ventilation and continued to maintain SpO2 of 98-100%. In view of the continued picture of hypoxaemia on serial ABGs, normal SpO2 and presence of blast crisis, a diagnosis of pseudohypoxaemia was made. Since point-of-care ABG analyser was not available, ABG repeated as rapidly as possible (delay of 7 minutes) from the sample immediately stored in ice showed a PaO2 of 71 mm Hg on FiO2 of 0.4 which was better than about 50 mm Hg consistently obtained in all the previous reports. We were prompted to precool the syringe even before obtaining the sample so that cooling the sampled blood could be hastened and the metabolism of cells be reduced more promptly.
tored in ice showed a PaO2 of 71 mm Hg on FiO2 of 0.4 which was better than about 50 mm Hg consistently obtained in all the previous reports. We were prompted to precool the syringe even before obtaining the sample so that cooling the sampled blood could be hastened and the metabolism of cells be reduced more promptly. We did three ABGs, one from sample obtained in a precooled syringe (kept in the freezer for 30 minutes and taken out just before drawing blood from a pre-existing arterial line), second from syringe immediately cooled in ice after sampling and another sample stored similarly but after a time delay of about 7 minutes after withdrawal. The time delay between sampling and analysis in first two samples was about 3 minutes. The ABGs are depicted in Table 3. Table 3 Serial ABG analysis of samples obtained from precooled syringe and without precooling These values not only proved the diagnosis of pseudohypoxaemia but also indicated the usefulness of precooling of the syringe before sampling. We also noticed a minimal increase in PaCO2 values in samples analysed from the syringes without precooling. The patient was extubated, shifted out of ICU and recovered uneventfully in the ward and was discharged on hydroxyurea and allopurinol medications.
These values not only proved the diagnosis of pseudohypoxaemia but also indicated the usefulness of precooling of the syringe before sampling. We also noticed a minimal increase in PaCO2 values in samples analysed from the syringes without precooling. The patient was extubated, shifted out of ICU and recovered uneventfully in the ward and was discharged on hydroxyurea and allopurinol medications. Discussion Fox et al first described the term ‘leucocytic larceny’ in 1979 as false low plasma oxygen tension measurement in leukaemia due to increased oxygen consumption by WBCs.[1] WBCs are about 45 times more metabolically active than platelets. However, since the platelet number far exceeds the WBCs, both cell types would contribute equally to the amount of oxygen consumption.[2] Blast cells being more metabolically active, consume more oxygen.[3] It is also reported that they would coat the sensing electrodes resulting in interference with analysis.[4] Patients with leukaemia and myelodysplasia show false reduction in PaO2 values associated with time delay in analysis of the sample. Metabolically active cells in the sample continue to consume oxygen resulting in such a reduction which is widely regarded as spurious hypoxaemia or pseudohypoxaemia.
Discussion Fox et al first described the term ‘leucocytic larceny’ in 1979 as false low plasma oxygen tension measurement in leukaemia due to increased oxygen consumption by WBCs.[1] WBCs are about 45 times more metabolically active than platelets. However, since the platelet number far exceeds the WBCs, both cell types would contribute equally to the amount of oxygen consumption.[2] Blast cells being more metabolically active, consume more oxygen.[3] It is also reported that they would coat the sensing electrodes resulting in interference with analysis.[4] Patients with leukaemia and myelodysplasia show false reduction in PaO2 values associated with time delay in analysis of the sample. Metabolically active cells in the sample continue to consume oxygen resulting in such a reduction which is widely regarded as spurious hypoxaemia or pseudohypoxaemia. Various reports of spurious reduction in the PaO2 values without clinical/radiological evidence of a causative insult have been reported.[5–7] They have recommended use of potassium cyanide,[8] and sodium fluoride to rapidly inhibit oxygen consumption by cells in the sample, use of plasma instead of whole blood,[4] and cooling the specimen in ice immediately after sampling to overcome this problem.[37] We noticed that placing the sample in ice immediately after sampling still resulted in low PaO2 values as reported earlier.[8–10] For accurate results, rapid analysis of the sample is crucial in such circumstances which can be obtained using Bed-side point-of-care analysers and continuous blood gas analysis.[6] However, these gadgets are expensive and hence, alternate methods are deviced to overcome this problem.
values as reported earlier.[8–10] For accurate results, rapid analysis of the sample is crucial in such circumstances which can be obtained using Bed-side point-of-care analysers and continuous blood gas analysis.[6] However, these gadgets are expensive and hence, alternate methods are deviced to overcome this problem. We noticed that precooling of the syringe prior to sampling helped reduce the oxygen consumption much more that cooling them after sampling. Precooling will definitely reduce the temperature of sampled blood more rapidly resulting in reliable PaO2 values. Pulse oximetry measures oxygen saturation of haemoglobin directly being unaffected by plasma oxygen tension. Hence, normal SpO2 values confirm pseudohypoxaemia in such conditions and also reliably guides oxygen therapy.[10] We suggest that the entity of spurious hypoxaemia should be familiar among all clinicians who treat patients with leukocytosis and thrombocytosis. All efforts should be made to minimise the time lag in analysing the blood gases and the values should be treated only after correlating clinically. Precooled syringes offer additional solution in efforts to rapidly reduce oxygen consumption by cells in the sampled blood. Pulse oximetry is most useful monitoring tool in diagnosing pseudohypoxaemia and guiding oxygen therapy. Medical management should aim at rapid cytoreduction in blast crisis which would help resolve these management issues. Source of Support: Nil Conflict of Interest: None declared.